# CT scanner: Interventions

Source page: https://www.materialinsights.org/products/p_ct_scanner/prod-interventions/

#### Interventions

Changes that would cut emissions or material use, by lever and life stage, with effect per unit, readiness, adoption, evidence and who acts

| Lever | Life stage | Change | Effect per unit | Readiness | Adoption | Who acts | Source |
|---|---|---|---|---|---|---|---|
| Efficiency | Use | Put the scanner in Off or low-power mode overnight instead of idle | COCIR's exemplary values are 72 kWh per day idle and 37 kWh per day Off; the 35 kWh difference (about 49%) is calculated ^[11] | Deployed (scenarios defined in the COCIR scheme) | Not stated | Hospitals and imaging centres | ^[11] |
| Efficiency | Use | Use an energy-savings mode | GE HealthCare says Energy Savings Mode cuts energy use by up to 15% on the Revolution Maxima ^[13] | Deployed | Not stated | Makers and hospitals | ^[13] |
| Design | Components design | Drop lead counterweights and cut lead in shielding | Siemens removed lead counterweights and says it cut lead shielding while protection stays equal ^[3]; GE HealthCare uses steel as counterweight ^[13]; no mass figure found | Deployed | Not stated | Makers | ^[3]^[13] |
| Circularity | End of life | Refurbish X-ray tube assemblies and reuse their parts | Up to 40% of a Siemens tube assembly may consist of reused parts; GE HealthCare reuses 60.1% of the mass of its tubes ^[3]^[13] | Deployed | Siemens takes tube assemblies back at end of life; no volume found | Makers | ^[3]^[13] |
| Circularity | Making | Recycle gadolinium oxide in detector ceramic production | About 25% of the gadolinium oxide processed each year comes from the closed loop, and the oxide supplied per area fell 82% against 2000/2001 ^[3] | Deployed | About 25% of annually processed gadolinium oxide ^[3] | Makers | ^[3] |
| Business model | End of life | Refurbish and resell whole scanners | Siemens refurbishes systems and reuses components and replacement parts where possible ^[3]; a sector report puts the refurbishment potential of CT at 85% (Table 6 value, not defined) ^[10]; no energy or emissions saving per unit was confirmed | Deployed (Siemens Lifecycle Product business ^[3]) | Not stated | Makers, service firms, hospitals | ^[3]^[10] |

as of 2026-11-14

#### Sources

1. [SOMATOM Pro.Pulse · Environmental Product Declaration (17893 0426)](https://marketing.webassets.siemens-healthineers.com/0e0e7165ae62df95/db61ce8b9818/CT-SOMATOM-Pro.Pulse-Environment-Product-Declaration.PDF) Siemens Healthineers AG, 2026, 'Climate change: total' chart; Key environmental performance indicators table
2. [Medical Imaging Equipment: State of Sustainability Research (final, 16 June 2022)](https://globalelectronicscouncil.org/medical-imaging-equipment-state-of-sustainability-research) Global Electronics Council for the Medical Equipment Proactive Alliance (MEPA), 2022, Tables 2 and 6
3. [COCIR Self-Regulatory Initiative for the Ecodesign of Medical Imaging Equipment: Status Report 2018](https://www.cocir.org/wp-content/uploads/COCIR_SRI_Status_Report_2018_-_June_2019.pdf) COCIR, 2018, Section 4.3.1
4. [Revolution Maxima: Focusing on sustainability in computed tomography (JB19858XX)](https://events.dev.gehealthcare.com/wp-content/uploads/2024/12/brochure_revolution_maxima_environmental_collateral_JB19858XX.pdf) GE HealthCare, 2024

## Other sections

- [Summary](https://www.materialinsights.org/products/p_ct_scanner/index.md)
- [Working](https://www.materialinsights.org/products/p_ct_scanner/prod-working/index.md)
- [Bill of materials](https://www.materialinsights.org/products/p_ct_scanner/prod-bom/index.md)
- [Use case](https://www.materialinsights.org/products/p_ct_scanner/prod-use/index.md)
- [Supply chain](https://www.materialinsights.org/products/p_ct_scanner/prod-supply/index.md)
- [Footprint](https://www.materialinsights.org/products/p_ct_scanner/prod-footprint/index.md)
- [Evolution](https://www.materialinsights.org/products/p_ct_scanner/prod-evolution/index.md)