# CT scanner: Working

Source page: https://www.materialinsights.org/products/p_ct_scanner/prod-working/

#### What it is

The product and the job it does, in a short paragraph

A CT scanner is a ring-shaped machine that turns an X-ray tube and a ring of detectors around a patient on a moving table and has a computer build slice images of the inside of the body. The functional unit is one complete medical CT scanner: gantry with tube and detectors, patient table, generator, cooling and operator console. Industrial and security CT, tubes and detectors sold alone, and plain X-ray machines (their own node) are left out.

as of 2026-11-14

#### How it works

How it does its job, step by step in plain words

The tube and the detectors sit on a rotating ring inside the gantry. The newest scanners spin it in as little as 0.33 s, and the table slides the patient through the ring ^[5]. Early scanners used a thin pencil beam, widened into a small fan in the second generation; in the third generation the tube and a detector array turn together around the patient, and that geometry is used in all commercial scanners today ^[1]. Detectors of the common kind turn X-rays into light in a ceramic and then into a signal. Photon-counting detectors convert each X-ray photon directly into an electrical signal ^[7]. Less than 1% of the tube's electron energy leaves as X-rays and more than 99% is heat, so tubes need strong cooling ^[30]. The computer turns thousands of readings into slices. The twin technology page explains the reconstruction.

as of 2026-11-14

#### Key specifications

The numbers that describe it, each with unit, variant, year and source

| Spec | Value | Unit | Variant | Year | Source |
|---|---|---|---|---|---|
| Weight of a SOMATOM Pro.Pulse CT scanner (Siemens declaration) | 2700 | kg (at most) | Conventional | 2026 | ^[3] |
| Weight of a NAEOTOM Alpha.Pro CT scanner (Siemens declaration) | 3300 | kg (approx.) | Photon-counting | 2025 | ^[4] |
| Fastest gantry rotation time (SOMATOM Pro.Pulse; also 0.5 s and 1.0 s) | 0.33 | s | Conventional | 2026 | ^[5] |
| Temporal resolution (SOMATOM Pro.Pulse) | 86 | ms | Conventional | 2026 | ^[5] |
| Temporal resolution, best case (NAEOTOM Alpha.Pro) | 66 | ms (down to) | Photon-counting | 2026 | ^[6] |
| Generator power (SOMATOM Pro.Pulse, 2 x 75 kW) | 150 | kW | Conventional | 2026 | ^[5] |
| Power of each of the two tubes (NAEOTOM Alpha.Pro) | 120 | kW per tube | Photon-counting | 2026 | ^[6] |
| Maximum power load (SOMATOM Pro.Pulse) | 200 | kVA (at most) | Conventional | 2026 | ^[3] |
| Maximum power load (NAEOTOM Alpha.Pro) | 350 | kVA (at most) | Photon-counting | 2025 | ^[4] |
| In-plane spatial resolution in Quantum HD mode (NAEOTOM Alpha.Pro) | 0.11 | mm | Photon-counting | 2026 | ^[6] |

as of 2026-11-14

#### Variants

The variants whose make-up or footprint differs a lot, what defines each and its share of units

| Variant | Defining spec | Year | Source |
|---|---|---|---|
| Conventional | Scintillator detector: a gadolinium oxysulfide ceramic (UFC) turns X-rays into light first ^[9]; reference SOMATOM Pro.Pulse with two tubes and two detectors ^[5] | 2026 | ^[3]^[5]^[9] |
| Photon-counting | Photon-counting detector that converts each X-ray photon directly into an electrical signal; the first such scanner, the NAEOTOM Alpha, has a cadmium telluride layer ^[7]. Reference NAEOTOM Alpha.Pro with two tubes and two QuantaMax detectors ^[6], which Siemens describes as crystal semiconductors ^[32]; its declaration names no detector material ^[4] | 2025 | ^[4]^[6]^[7]^[32] |

Indicative · as of 2026-11-14. Each variant is one Siemens model, the only ones with a published life-cycle declaration. Both are dual-source scanners, so weight and footprint also reflect that design. No source gives shares of units sold, and none opened names the detector material of the NAEOTOM Alpha.Pro itself.

#### Limits

The physical limits set by science and the engineering limits of today's design, with today's value

| Limit | Kind | Today | Unit | Source |
|---|---|---|---|---|
| Share of the tube's electron energy that leaves as X-rays; more than 99% is heat | Physical | 1 | % (less than) | ^[30] |
| Shortest time needed for one image, which sets how well a beating heart is frozen (best case today) | Engineering | 66 | ms | ^[6] |
| Finest detail resolved in the image plane in the best mode of today's photon-counting scanner | Engineering | 0.11 | mm | ^[6] |

as of 2026-11-14

#### Science it relies on

The science principles it works by or is limited by
 Relies on

- [X-rays](https://www.materialinsights.org/science/xrays/) Dense tissue absorbs more X-rays, so the beam that gets through carries the picture.
- [Fourier analysis](https://www.materialinsights.org/science/fourier_analysis/) Rebuilding a slice from many line measurements is a mathematical problem; the twin technology page explains it.
 Limited by

- [X-rays](https://www.materialinsights.org/science/xrays/) X-rays ionise tissue, so EU rules require each dose to fit the medical purpose of the exposure ^[26]; regulators also license the equipment ^[17].

as of 2026-11-14

#### Sources

1. [CT Archives - AAPM Virtual Museum](https://museum.aapm.org/exhibit/07-ct) AAPM Virtual Museum (American Association of Physicists in Medicine)
2. [SOMATOM Pro.Pulse · Environmental Product Declaration (17893 0426)](https://marketing.webassets.siemens-healthineers.com/0e0e7165ae62df95/db61ce8b9818/CT-SOMATOM-Pro.Pulse-Environment-Product-Declaration.PDF) Siemens Healthineers AG, 2026, 'Climate change: total' chart; Key environmental performance indicators table
3. [NAEOTOM Alpha.Pro · Environmental Product Declaration (17247 1125)](https://marketing.webassets.siemens-healthineers.com/c87c68811e34fdb4/28c80a4d5d05/siemens-healthineers-CT_NAEOTOM_Alpha_Pro_EPD.pdf) Siemens Healthineers AG, 2025, 'Climate change: total' chart; Key environmental performance indicators table
4. [SOMATOM Pro.Pulse Dual Source CT scanner](https://www.siemens-healthineers.com/computed-tomography/somatom/somatom-propulse) Siemens Healthineers, 2026
5. [NAEOTOM Alpha.Pro](https://www.siemens-healthineers.com/en-us/computed-tomography/naeotom/naeotom-alpha-pro) Siemens Medical Solutions USA, 2026
6. [Siemens Healthineers Announces FDA Clearance of NAEOTOM Alpha, World's First Photon-Counting CT](https://www.siemens-healthineers.com/en-us/press-room/press-releases/fdaclearacenaeotomalpa) Siemens Healthineers, 2021
7. [UFC - Ultra Fast Ceramic](https://www.siemens-healthineers.com/si/computed-tomography/ufc) Siemens Healthineers, 2026
8. [AERB Annual Report 2023, Chapter 2: Safety Surveillance of Radiation Facilities](https://mgw3.aerb.gov.in/images/PDF/Annual_report/ar2023/chap/CHAPTER-02.pdf) Atomic Energy Regulatory Board, India, 2023, Tables 2.1 and 2.2
9. [Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation](https://www.legislation.gov.uk/eudr/2013/59/article/5/data.htm) legislation.gov.uk (EU text as it stood on 31 December 2020), 2013, Article 5
10. [X-ray production](https://radiopaedia.org/articles/25428) Radiopaedia.org
11. [Photon-counting CT by Siemens Healthineers: Quantum Technology](https://www.siemens-healthineers.com/computed-tomography/ct-technologies-and-innovations/photon-counting-ct) Siemens Healthineers, 2026

## Other sections

- [Summary](https://www.materialinsights.org/products/p_ct_scanner/index.md)
- [Bill of materials](https://www.materialinsights.org/products/p_ct_scanner/prod-bom/index.md)
- [Use case](https://www.materialinsights.org/products/p_ct_scanner/prod-use/index.md)
- [Supply chain](https://www.materialinsights.org/products/p_ct_scanner/prod-supply/index.md)
- [Footprint](https://www.materialinsights.org/products/p_ct_scanner/prod-footprint/index.md)
- [Evolution](https://www.materialinsights.org/products/p_ct_scanner/prod-evolution/index.md)
- [Interventions](https://www.materialinsights.org/products/p_ct_scanner/prod-interventions/index.md)