#### What it is

The product and the job it does, in a short paragraph

An X-ray apparatus is a machine that sends a controlled beam of X-rays through the body and records the shadow on a detector, so that doctors and dentists can see bones, lungs and teeth. The functional unit is one complete diagnostic X-ray machine, from generator and tube to stand or table, detector and console. CT scanners have their own node; X-ray tubes sold alone, security and industrial X-ray machines and radiotherapy are left out. ^[3]

as of 2026-11-14

#### How it works

How it does its job, step by step in plain words

A high-voltage generator feeds an X-ray tube. Inside the tube a heated tungsten filament gives off electrons in a high vacuum, and they strike a metal target ^[14]^[15]. Less than 1% of the electrons' energy comes out as X-rays and more than 99% becomes heat ^[16]. A collimator shapes the beam, which passes through the patient. The flat-panel detectors now common in radiography turn the X-rays that get through into light and then into an electrical signal, using a phosphor such as caesium iodide ^[13]. A computer in the console processes the image at once, with no film to develop. Mobile units and C-arms use the same chain on wheels or on a rotating arm. The working principle and its history are on the twin technology page.

as of 2026-11-14

#### Key specifications

The numbers that describe it, each with unit, variant, year and source

| Spec | Value | Unit | Year | Source |
|---|---|---|---|---|
| Weight of one Siemens MULTIX Impact radiography system | 1184 | kg | 2022 | ^[3] |
| Power drawn with the system in operation (MULTIX Impact) | 392 | W | 2022 | ^[3] |
| Power drawn with the system in standby (MULTIX Impact) | 340 | W | 2022 | ^[3] |
| Power drawn with the system off and detectors in standby (MULTIX Impact) | 110 | W | 2022 | ^[3] |
| Daily energy use of the MULTIX Impact under the declaration's daily profile (see usage pattern) | 5.26 | kWh per day | 2022 | ^[3] |
| Continuous power in scan mode (GE HealthCare Definium Pace Select) | 2.2 | kVA | 2023 | ^[11] |
| Power in standby (GE HealthCare Definium Pace Select) | 0.85 | kVA | 2023 | ^[11] |
| Share of daily energy used in scan mode (COCIR guidance for X-ray equipment) | 3 | % (around) | 2015 | ^[5] |

as of 2026-11-14

#### Limits

The physical limits set by science and the engineering limits of today's design, with today's value

| Limit | Kind | Today | Unit | Source |
|---|---|---|---|---|
| Share of the electron beam's energy that leaves the tube as X-rays; more than 99% is heat the tube must shed | Physical | 1 | % (less than) | ^[16] |
| Refurbishment potential of an X-ray unit (MEPA Table 6 value; the report does not define the term) | Engineering | 60 | % (refurbishment potential) | ^[4] |

as of 2026-11-14

#### Science it relies on

The science principles it works by or is limited by
 Relies on

- [X-rays](https://www.materialinsights.org/science/xrays/) The image is a shadow made by tissue that absorbs X-rays unequally.
- [Thermionic emission](https://www.materialinsights.org/science/thermionic_emission/) The tube's electrons come from a heated tungsten filament in a high vacuum ^[14].
 Limited by

- [X-rays](https://www.materialinsights.org/science/xrays/) X-rays ionise tissue, so EU rules require each dose to fit the medical purpose of the exposure, which caps how much radiation an image may use ^[21]; regulators also license the equipment ^[7].

as of 2026-11-14

#### Sources

1. [MULTIX Impact | Environmental Product Declaration](https://storage.mtender.gov.md/get/d0c165f0-027a-426e-bc8b-fcbefcd19fa1-1724056431740) Siemens Healthcare GmbH (copy held on the Moldovan public procurement portal MTender; no copy found on Siemens Healthineers' own site), 2022
2. [Medical Imaging Equipment: State of Sustainability Research (final, 16 June 2022)](https://globalelectronicscouncil.org/medical-imaging-equipment-state-of-sustainability-research) Global Electronics Council for the Medical Equipment Proactive Alliance (MEPA), 2022, Table 6 and section 2
3. [COCIR Guidelines for Users on Saving Energy: Good Environmental Practice (X-ray, October 2015)](https://dev.cocir.org/wp-content/uploads/COCIR_Guidelines_for_users_for_saving_energy_-_X-ray_Sept2015.pdf) COCIR, 2015
4. [AERB Annual Report 2023, Chapter 2: Safety Surveillance of Radiation Facilities](https://mgw3.aerb.gov.in/images/PDF/Annual_report/ar2023/chap/CHAPTER-02.pdf) Atomic Energy Regulatory Board, India, 2023, Tables 2.1 and 2.2
5. [Focusing on sustainability in X-ray solutions: Definium Pace Select (JB25106XX)](https://Www.Gehealthcare.Com/-/jssmedia/gehc/us/files/about-us/sustainability/reports/brochure-environmental-collateral-definiumpace-jb25106xx.pdf?rev=-1) GE HealthCare, 2023
6. [Digital radiography: The bottom line comparison of CR and DR technology](https://www.appliedradiology.com/articles/digital-radiography-the-bottom-line-comparison-of-cr-and-dr-technology) Applied Radiology (J. Anthony Seibert), 2009
7. [WE-H-204-01: William D. Coolidge, Inventor of the Modern X-Ray Tube](https://www.osti.gov/biblio/22679129) Medical Physics, vol. 43 no. 6 (abstract by David J. Allard), record on OSTI.gov, 2016
8. [William D. Coolidge](https://edisontechcenter.org/coolidge.html) Edison Tech Center
9. [X-ray production](https://radiopaedia.org/articles/25428) Radiopaedia.org
10. [Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation](https://www.legislation.gov.uk/eudr/2013/59/article/5/data.htm) legislation.gov.uk (EU text as it stood on 31 December 2020), 2013, Article 5

## Other sections

- [Summary](https://www.materialinsights.org/products/p_x_ray_apparatus/index.md)
- [Bill of materials](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-bom/index.md)
- [Use case](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-use/index.md)
- [Supply chain](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-supply/index.md)
- [Footprint](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-footprint/index.md)
- [Evolution](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-evolution/index.md)
- [Interventions](https://www.materialinsights.org/products/p_x_ray_apparatus/prod-interventions/index.md)