# Magnetic resonance imaging: Levers

Source page: https://www.materialinsights.org/technologies/mri_scanner/tech-levers/

Held back most by

Economics

The physics and the magnet design are mature; what holds India back is the price of a scanner (about Rs 8 crore, per an AIIMS official in 2025 ^[11]), which an indigenous 1.5 tesla machine aims to cut by up to 40 to 50 per cent ^[11]. Helium is the world-scale limit to watch: MRI took 15 per cent of US helium use in 2025 ^[4], but sealed magnets cut the helium held from 1,500 litres to 7 litres in the maker's own claim ^[10] (judgement on which binds first).

#### The six layers

Mathematics, science, engineering, production, economics, and resources and rules: what each gives, the idea in plain words, the limit (hard or soft) and how close today is, what could unlock it, and the maths behind it

Six layers make a technology possible, from the mathematics underneath to the resources and rules it needs at world scale. A **hard limit** comes from mathematics or nature and can only be sidestepped; a **soft limit** moves with effort, scale, money or policy.

1.  1 Mathematics Hard limit

Mathematics is the technology

**What it gives.** The step that turns a faint radio signal into a picture: a Fourier transform sorts the signal by frequency, and frequency stands for position.

**The idea.** Role: foundation. The Larmor law is physics, but the picture itself is a mathematical construction: a field gradient makes each position ring at its own frequency ^[15], and the 2D Fourier transform reconstructs an image from the sampled data ^[9]. Without that step the scanner has a signal but no image, as with a CT scanner (judgement).

**The limit.** The picture is only as detailed as the number of samples collected, and each sample costs scan time, so detail trades against time; this is a limit of sampling (judgement).

**What could unlock it.** Smarter sampling and reconstruction can fill in missing data from fewer measurements, which shortens scans (judgement).

[Fourier analysis](https://www.materialinsights.org/science/fourier_analysis/)[AIx AI diagnosis from medical images](https://www.materialinsights.org/technologies/ai_medical_imaging/)

Show the maths ω(x) = γ (B0 + Gx x)

**ω(x)**
: how fast the nuclei at position x rotate, in radians a second

**γ**
: gyromagnetic ratio; for hydrogen 2π × 42.58 MHz per tesla, about 2.675 × 10^8 radians a second per tesla (calculated from ^[8])

**B0**
: the strong steady field of the magnet, in tesla

**Gx**
: how fast the field is made to change along the direction x, in tesla per metre

**x**
: position along that direction, in metres

**In plain words.** The magnet is made a little stronger at one side of the body than the other. So the nuclei at each place spin at their own rate and send a signal with their own pitch. The computer splits the mixed signal into its pitches, using a Fourier transform. Each pitch tells it a position, and that builds the picture.

**Worked example.** With a gradient of 10 mT per metre (assumed), a point 10 cm from the centre is shifted by 42.58 MHz per tesla × 0.010 T per metre × 0.1 m = 42.58 kHz (calculated from ^[8]^[15]), so two points 1 mm apart differ by about 426 Hz (calculated).

2.  2 Science Hard limit

**What it gives.** A signal worth detecting: hydrogen nuclei in water respond to radio waves at a frequency fixed by the magnetic field.

**The idea.** Nuclei in a strong magnetic field rotate at a frequency that depends on the strength of the field ^[1]. For hydrogen the frequency is 42.58 MHz for each tesla of field ^[8], so a 1.5 T scanner works near 63.87 MHz and a 3 T scanner near 127.74 MHz (calculated from ^[8]).

**The limit.** The scanner can only listen at the frequency its field sets (the Larmor frequency), so signal and detail rest on the size and steadiness of the field; no ionising radiation is used ^[1] (judgement on the first clause).

**What could unlock it.** Stronger magnets give more signal but need bigger magnets and more helium cooling; the same physics runs in chemical NMR labs ^[8].

[Nuclear magnetic resonance](https://www.materialinsights.org/science/nuclear_magnetic_resonance/)[Superconductivity](https://www.materialinsights.org/science/superconductivity/)

Show the maths f = γ̄ B0

**f**
: the Larmor frequency, how many times a second the nuclei rotate, in megahertz

**γ̄**
: gyromagnetic ratio divided by 2π, a fixed number for each kind of nucleus; for hydrogen it is 42.58 MHz per tesla ^[8]

**B0**
: the strength of the magnet's field, in tesla

**In plain words.** Hydrogen nuclei act like tiny spinning magnets. In a strong field they wobble at a speed set by that field. Double the field and they wobble twice as fast. The scanner sends radio waves at exactly that speed to talk to them.

**Worked example.** At 3 T the frequency is 42.58 × 3 = 127.74 MHz (calculated from ^[8]); the Martinos Center slides print about 127.7 MHz ^[9].

3.  3 Engineering Soft limit

**What it gives.** A magnet that is strong, steady over the whole body and cooled to a few degrees above absolute zero.

**The idea.** Most clinical scanners make their field with superconducting coils, which carry current with no resistance once cooled with liquid helium (judgement, as in the node's science tab). A conventional magnet holds 1,500 litres of helium; one maker's sealed design holds 7 litres ^[10].

**The limit.** Field strength, steadiness across the body and cooling set the cost and size of the magnet; the 7 litre figure is a maker's claim, not an independent test ^[10] (judgement).

**What could unlock it.** Sealed magnets with little helium, and in India an indigenous 1.5 tesla magnet being built by IUAC ^[11].

[He Helium](https://www.materialinsights.org/commodities/helium/)[Superconductivity](https://www.materialinsights.org/science/superconductivity/)

4.  4 Production Soft limit

**What it gives.** Many working scanners from a few suppliers of magnets, coils and electronics.

**The idea.** Even India's indigenous machine still relies on a few imported components, such as superconducting wire for the magnet, according to a 2025 report ^[11]. Under the production-linked incentive scheme, production has started for 54 products including MRI, ultrasound and CT machines ^[12].

**The limit.** Superconducting wire, coils and the know-how to make a magnet uniform are held by few firms, so a new maker must build all of this (judgement).

**What could unlock it.** A government-funded indigenous scanner developed by SAMEER with CDAC and IUAC and seven to eight industry partners, fully funded by MeitY ^[11], and the incentive scheme ^[12].

[MRI scanner](https://www.materialinsights.org/products/p_mri_scanner/)

5.  5 Economics Soft limit Binds most today

**What it gives.** A price that a hospital can recover from scans.

**The idea.** An MRI machine costs around Rs 8 crore, according to a senior AIIMS official quoted in 2025; the indigenous machine promises to cut the cost of this equipment by up to 40 to 50 per cent ^[11]. If it did, Rs 8 crore would fall to Rs 4 to 4.8 crore (calculated from ^[11]).

**The limit.** The price of the machine, its helium and servicing, and enough patients to pay for scans (judgement).

**What could unlock it.** A cheaper domestic machine and a bigger local supply chain ^[11], and the incentive scheme, whose cumulative eligible sales reached Rs 10,413.40 crore up to March 2025 across all device types ^[12].

6.  6 Resources and rules Soft limit

**What it gives.** Permission to sell, and enough helium to cool the world's magnets.

**The idea.** MRI took 15 per cent of helium use in the United States in 2025, according to the US Geological Survey ^[4]. In India all medical devices are regulated under the Medical Devices Rules, 2017 since 11 February 2020 ^[14], and CDSCO's 2019 notice lists MRI as Class C ^[13], the moderate high risk class ^[14], which is licensed by the Central Licensing Authority ^[14].

**The limit.** Helium is extracted from natural gas ^[4], so MRI competes for a gas that is hard to scale quickly, and every unit needs a Class C approval ^[13]^[14] (judgement on the scaling).

**What could unlock it.** Sealed low-helium magnets cut the amount needed per scanner from 1,500 litres to 7 litres, a ratio of about 214 to 1 (calculated from ^[10]).

[He Helium](https://www.materialinsights.org/commodities/helium/)

as of 2026-11-25

#### Sources

1. [The Nobel Prize in Physiology or Medicine 2003: press release (6 October 2003)](https://nobelprize.org/prizes/medicine/2003/press-release) Nobel Assembly at Karolinska Institutet (nobelprize.org), 2003
2. [Mineral Commodity Summaries 2026: Helium](https://pubs.usgs.gov/periodicals/mcs2026/mcs2026-helium.pdf) US Geological Survey, 2026
3. [Magnetic Resonance Imaging, Lecture 1 Section 1 (Physics of Medical Imaging and Radiotherapy)](https://ccap.hep.ph.imperial.ac.uk/trac/raw-attachment/wiki/Teaching/2020-22/PhMIRT/MRI-Lctr01-Sctn01.pdf) Imperial College London, Department of Physics (K. Long), 2022
4. [Why and How: MRI Physics](https://gate.nmr.mgh.harvard.edu/wiki/whynhow/images/3/35/20191010_McDaniel_MRIPhysics.pdf) Patrick McDaniel, Massachusetts General Hospital (Martinos Center), 2019
5. [Philips extends leadership in virtually helium free MRI with more than 1,000 systems installed (company press release, 28 February 2024)](https://www.usa.philips.com/about/news-and-insights/press-release/philips-extends-leadership-in-virtually-helium-free-mri-with-more-than-1000-systems-installed-expanding-patient-access-worldwide) Philips, 2024
6. [Story behind India's first indigenous MRI machine that could make scans more affordable (28 March 2025)](https://theprint.in/health/story-behind-indias-1st-indigenous-mri-machine-that-could-make-scans-more-affordable/2568798) ThePrint, 2025
7. [Lok Sabha Unstarred Question 2126, answered 1 August 2025 (PLI scheme for medical devices)](https://eparlib.sansad.in/bitstream/123456789/2993872/1/AU2126_oR2aGu.pdf) Department of Pharmaceuticals, Ministry of Chemicals and Fertilizers, Government of India, 2025
8. [Medical device classification list, notice of 15 May 2019](https://cdsco.gov.in/opencms/resources/UploadCDSCOWeb/2018/UploadPublic_NoticesFiles/MD%20final%20classification%20list%20merged.pdf) Central Drugs Standard Control Organization (CDSCO), 2019
9. [Frequently Asked Questions on Medical Devices Rules, 2017 (CDSCO/FAQ/MD/01/2024, year from document number)](https://cdsco.gov.in/opencms/export/sites/CDSCO_WEB/Pdf-documents/mdfaq24.pdf) Central Drugs Standard Control Organization (CDSCO), 2024
10. [Phase Encoding: Gradients and Phase Shift (course notes, undated)](https://watkins.cs.queensu.ca/~jstewart/472/notes/08/08i-phase-encoding.html) Queen's University, School of Computing (CISC 472 notes), 2026

## Other sections

- [Summary](https://www.materialinsights.org/technologies/mri_scanner/index.md)
- [Overview](https://www.materialinsights.org/technologies/mri_scanner/tech-overview/index.md)
- [Science](https://www.materialinsights.org/technologies/mri_scanner/tech-science/index.md)
- [Evolution](https://www.materialinsights.org/technologies/mri_scanner/tech-evolution/index.md)
- [Materials](https://www.materialinsights.org/technologies/mri_scanner/tech-materials/index.md)
- [Energy footprint](https://www.materialinsights.org/technologies/mri_scanner/tech-energy-footprint/index.md)
- [Industries](https://www.materialinsights.org/technologies/mri_scanner/tech-industries/index.md)