MRI scanner: Working
What it is
The product and the job it does, in a short paragraph
An MRI scanner takes detailed pictures of the inside of the body, mostly soft tissue such as the brain, spine, joints and organs. It uses a strong magnet, radio waves and a computer, and no ionising radiation [9]. The functional unit is one whole-body scanner for medical diagnosis, of any field strength, with its magnet, coils and computers. Contrast agents, small research scanners and portable scanners are left out.
as of 2026-11-13
How it works
How it does its job, step by step in plain words
The patient lies inside a large magnet that gives a strong, steady field. In most clinical scanners the magnet is a coil of superconducting wire, kept cold by liquid helium so that it carries current without loss [5][18]. Gradient coils vary the field across the body and radio-frequency coils send and receive the signals. A computer turns the signals into images. A small refrigerator called a cold head helps keep the magnet cold, and Siemens lists its replacement interval as 2 years in one declaration's table [1]. Conventional magnets hold about 1,500 litres of liquid helium; sealed designs keep a small charge for the life of the magnet [4][5]. The physics is set out in the twin technology page.
as of 2026-11-13
Key specifications
The numbers that describe it, each with unit, variant, year and source
| Spec | Value | Unit | Variant | Year | Source |
|---|---|---|---|---|---|
| Total weight of MAGNETOM Amira (1.5T), Siemens declaration | 6416 | kg | 1.5T | 2016 | [3] |
| Total weight of MAGNETOM Vida (3T), Siemens declaration | 10,000 | kg | 3T | 2019 | [2] |
| Total weight of MAGNETOM Sola (1.5T), Siemens declaration | 6500 | kg | 1.5T | 2020 | [1] |
| Liquid helium in a conventional magnet, about (Philips: classic magnet; Siemens: as much as) | 1500 | litres | – | 2024 | [4][5][6] |
| Helium sealed in a Philips BlueSeal magnet | 7 | litres | – | 2024 | [4] |
| Liquid helium in a Siemens DryCool magnet (MAGNETOM Flow) for the life of the device | 0.7 | litres | – | 2024 | [5] |
| Power drawn, system off (MAGNETOM Sola, typical values at 400 V, 50 Hz; table value) | 4.3 | kW | 1.5T | 2020 | [1] |
| Power drawn, ready to measure (MAGNETOM Sola, XJ gradients; table value) | 8.2 | kW | 1.5T | 2020 | [1] |
| Power drawn, scanning (MAGNETOM Sola, XJ gradients; table value) | 20.2 | kW | 1.5T | 2020 | [1] |
| Power drawn, scanning (MAGNETOM Vida, XQ gradients; table value) | 23.1 | kW | 3T | 2019 | [2] |
| Primary energy demand over the life of MAGNETOM Sola, calculated as the sum of the declaration's table values: material supply 134, production and delivery 456, use and maintenance 2,874 (10 years) and end of life -55 MWh | 3409 | MWh primary energy | 1.5T | 2020 | [1] |
as of 2026-11-13
Variants
The variants whose make-up or footprint differs a lot, what defines each and its share of units
| Variant | Defining spec | Year | Source |
|---|---|---|---|
| 1.5T | Main field of 1.5 tesla; MAGNETOM Amira weighs about 6,416 kg and MAGNETOM Sola about 6,500 kg [3][1] | 2016 | [3][1] |
| 3T | Main field of 3 tesla; MAGNETOM Vida weighs about 10,000 kg, and non-ferrous metals are 40% of its weight against 30% in the 1.5T Amira (table values) [2][3] | 2019 | [2][3] |
as of 2026-11-13
Limits
The physical limits set by science and the engineering limits of today's design, with today's value
| Limit | Kind | Today | Unit | Source |
|---|---|---|---|---|
| Liquid helium needed to cool the superconducting magnet: conventional about 1,500 litres, sealed 7 litres (Philips) or 0.7 litres (Siemens) | Engineering | 0.7 | litres of helium per magnet (lowest found; no floor stated) | [4][5] |
| Allowed ambient temperature and relative humidity in the examination room (MAGNETOM Sola, table values) | Engineering | – | 18 to 22 degrees C, relative humidity 40 to 60% | [1] |
as of 2026-11-13
Science it relies on
The science principles it works by or is limited by
Relies on- Nuclear magnetic resonance Hydrogen nuclei in the body answer radio waves at a frequency set by the magnetic field; the scanner reads that answer.
- Superconductivity Clinical magnets use niobium-titanium wire that carries current with no resistance when cold; tens of thousands of such magnets have been built [18].
as of 2026-11-13
Sources
- MAGNETOM Sola Environmental Product Declaration Siemens Healthcare GmbH, 2020, product materials, power consumption, cumulative energy demand, life cycle of replacement parts and operating environment tables; recyclability text; 1.5T system
- MAGNETOM Vida Environmental Product Declaration Siemens Healthcare GmbH, 2019, product materials and power consumption tables; recyclability text; 3T system
- MAGNETOM Amira Environmental Product Declaration Siemens Healthcare GmbH, 2016, identification of product materials, power consumption and cumulative energy demand tables; recyclability text; 1.5T system
- Philips extends leadership in virtually helium-free MRI with more than 1,000 systems installed Philips, 2024, press release of 28 February 2024
- Siemens Healthineers introduces MAGNETOM Flow for greater sustainability and efficiency in MRI Siemens Healthineers, 2024, page dated 12 March 2024
- MRI's Helium Footprint – And How to Make it Harder to Spot Siemens Healthineers (Philipp Grätzel von Grätz, M.D.), article; the page shows no publication date
- The Nobel Prize in Physiology or Medicine 2003: press release (6 October 2003) Nobel Assembly at Karolinska Institutet (nobelprize.org), 2003, press release of 6 October 2003
- The Current Status and Future Direction of High-Magnetic-Field Science and Technology in the United States National Academies of Sciences, Engineering, and Medicine, 2024, Chapter 6, Superconducting Materials and Wires/Tapes for High-Field Magnets, pp. 146 and 151