MRI scanner: Working

What it is

The product and the job it does, in a short paragraph

An MRI scanner takes detailed pictures of the inside of the body, mostly soft tissue such as the brain, spine, joints and organs. It uses a strong magnet, radio waves and a computer, and no ionising radiation [9]. The functional unit is one whole-body scanner for medical diagnosis, of any field strength, with its magnet, coils and computers. Contrast agents, small research scanners and portable scanners are left out.

as of 2026-11-13

How it works

How it does its job, step by step in plain words

The patient lies inside a large magnet that gives a strong, steady field. In most clinical scanners the magnet is a coil of superconducting wire, kept cold by liquid helium so that it carries current without loss [5][18]. Gradient coils vary the field across the body and radio-frequency coils send and receive the signals. A computer turns the signals into images. A small refrigerator called a cold head helps keep the magnet cold, and Siemens lists its replacement interval as 2 years in one declaration's table [1]. Conventional magnets hold about 1,500 litres of liquid helium; sealed designs keep a small charge for the life of the magnet [4][5]. The physics is set out in the twin technology page.

as of 2026-11-13

Key specifications

The numbers that describe it, each with unit, variant, year and source

SpecValueUnitVariantYearSource
Total weight of MAGNETOM Amira (1.5T), Siemens declaration6416kg1.5T2016[3]
Total weight of MAGNETOM Vida (3T), Siemens declaration10,000kg3T2019[2]
Total weight of MAGNETOM Sola (1.5T), Siemens declaration6500kg1.5T2020[1]
Liquid helium in a conventional magnet, about (Philips: classic magnet; Siemens: as much as)1500litres– 2024[4][5][6]
Helium sealed in a Philips BlueSeal magnet7litres– 2024[4]
Liquid helium in a Siemens DryCool magnet (MAGNETOM Flow) for the life of the device0.7litres– 2024[5]
Power drawn, system off (MAGNETOM Sola, typical values at 400 V, 50 Hz; table value)4.3kW1.5T2020[1]
Power drawn, ready to measure (MAGNETOM Sola, XJ gradients; table value)8.2kW1.5T2020[1]
Power drawn, scanning (MAGNETOM Sola, XJ gradients; table value)20.2kW1.5T2020[1]
Power drawn, scanning (MAGNETOM Vida, XQ gradients; table value)23.1kW3T2019[2]
Primary energy demand over the life of MAGNETOM Sola, calculated as the sum of the declaration's table values: material supply 134, production and delivery 456, use and maintenance 2,874 (10 years) and end of life -55 MWh3409MWh primary energy1.5T2020[1]

as of 2026-11-13

Variants

The variants whose make-up or footprint differs a lot, what defines each and its share of units

VariantDefining specYearSource
1.5TMain field of 1.5 tesla; MAGNETOM Amira weighs about 6,416 kg and MAGNETOM Sola about 6,500 kg [3][1]2016[3][1]
3TMain field of 3 tesla; MAGNETOM Vida weighs about 10,000 kg, and non-ferrous metals are 40% of its weight against 30% in the 1.5T Amira (table values) [2][3]2019[2][3]

as of 2026-11-13

Limits

The physical limits set by science and the engineering limits of today's design, with today's value

LimitKindTodayUnitSource
Liquid helium needed to cool the superconducting magnet: conventional about 1,500 litres, sealed 7 litres (Philips) or 0.7 litres (Siemens)Engineering0.7litres of helium per magnet (lowest found; no floor stated)[4][5]
Allowed ambient temperature and relative humidity in the examination room (MAGNETOM Sola, table values)Engineering– 18 to 22 degrees C, relative humidity 40 to 60%[1]

as of 2026-11-13

Science it relies on

The science principles it works by or is limited by

Relies on
  • Nuclear magnetic resonance Hydrogen nuclei in the body answer radio waves at a frequency set by the magnetic field; the scanner reads that answer.
  • Superconductivity Clinical magnets use niobium-titanium wire that carries current with no resistance when cold; tens of thousands of such magnets have been built [18].

as of 2026-11-13

Sources

  1. MAGNETOM Sola Environmental Product Declaration Siemens Healthcare GmbH, 2020, product materials, power consumption, cumulative energy demand, life cycle of replacement parts and operating environment tables; recyclability text; 1.5T system
  2. MAGNETOM Vida Environmental Product Declaration Siemens Healthcare GmbH, 2019, product materials and power consumption tables; recyclability text; 3T system
  3. MAGNETOM Amira Environmental Product Declaration Siemens Healthcare GmbH, 2016, identification of product materials, power consumption and cumulative energy demand tables; recyclability text; 1.5T system
  4. Philips extends leadership in virtually helium-free MRI with more than 1,000 systems installed Philips, 2024, press release of 28 February 2024
  5. Siemens Healthineers introduces MAGNETOM Flow for greater sustainability and efficiency in MRI Siemens Healthineers, 2024, page dated 12 March 2024
  6. MRI's Helium Footprint – And How to Make it Harder to Spot Siemens Healthineers (Philipp Grätzel von Grätz, M.D.), article; the page shows no publication date
  7. The Nobel Prize in Physiology or Medicine 2003: press release (6 October 2003) Nobel Assembly at Karolinska Institutet (nobelprize.org), 2003, press release of 6 October 2003
  8. The Current Status and Future Direction of High-Magnetic-Field Science and Technology in the United States National Academies of Sciences, Engineering, and Medicine, 2024, Chapter 6, Superconducting Materials and Wires/Tapes for High-Field Magnets, pp. 146 and 151