Material Insights · Understanding the Material World

Stent: Working

What it is

The product and the job it does, in a short paragraph

A coronary stent is a small mesh tube that expands to hold a narrowed heart artery open; Medtronic describes it as made of medical-grade stainless steel or a cobalt alloy [4]. The functional unit is one stent, bare metal, drug-eluting or bioresorbable. HS 9021.90 also covers other implants. Stents for other vessels, stent grafts and the balloon catheter that delivers a stent are left out.

as of 2026-11-13

How it works

How it does its job, step by step in plain words

The stent sits on a small balloon that is carried into the coronary artery and inflated, pressing plaque back and restoring blood flow [4]. The mesh expands with the balloon, or in self-expanding designs opens by itself when a delivery system releases it [6]. Once the balloon is withdrawn the metal stent stays in the artery for good [4], and its struts hold the wall apart. Drug-eluting stents also carry a drug that slows tissue regrowth, held in a polymer coating on the metal [2][3]. Bioresorbable scaffolds are made to dissolve [6]. The working principle and history are on the twin technology page.

as of 2026-11-13

Key specifications

The numbers that describe it, each with unit, variant, year and source

SpecValueUnitVariantYearSource
Strut thickness of the Taxus Express stent (316L stainless steel, drug-eluting), as the 2010 review gives it132µmDrug-eluting2010[1]
Strut thickness of early-generation drug-eluting stents (stainless steel), low end (review figure)130µmDrug-eluting2017[2]
Strut thickness of early-generation drug-eluting stents (stainless steel), high end (review figure)150µmDrug-eluting2017[2]
Strut thickness of new-generation drug-eluting stents, low end (review figure)50µmDrug-eluting2017[2]
Strut thickness of new-generation drug-eluting stents, high end (review figure)90µmDrug-eluting2017[2]
Strut thickness of the Xience cobalt-chromium drug-eluting stent81µmDrug-eluting2024[6]
Strut thickness of the Absorb bioresorbable vascular scaffold150µmBioresorbable2024[6]
Lengths offered for every diameter of the XIENCE V stent, shortest8mmDrug-eluting2008[3]
Lengths offered for every diameter of the XIENCE V stent, longest28mmDrug-eluting2008[3]

as of 2026-11-13

Variants

The variants whose make-up or footprint differs a lot, what defines each and its share of units

VariantDefining specYearSource
Bare metalMetal mesh with no drug coating; India's ceiling price was INR 7,260 in February 2017 [9][23]; the government said bare metal stents had 10% of India's market then [23]2017[9][23]
Drug-elutingMetal mesh with a polymer layer that holds a drug; struts 130 to 150 µm in stainless steel stents, 50 to 90 µm in new-generation ones [2]; India's ceiling price was INR 29,600 in February 2017, when the government said drug-eluting stents had 90% of India's market [9][23]2017[2][9][23]
BioresorbableScaffold made to dissolve; the Absorb had struts of 150 µm against 81 µm for the Xience [6]; CE mark in January 2011 and FDA approval in July 2016 [7]2016[6][7]

as of 2026-11-13

Limits

The physical limits set by science and the engineering limits of today's design, with today's value

LimitKindTodayUnitSource
Strut thickness of the thinnest drug-eluting stent platform found (Xience, cobalt-chromium)Engineering81µm[6]
Strut thickness of the bioresorbable Absorb scaffoldEngineering150µm[6]

as of 2026-11-13

Science it relies on

The science principles it works by or is limited by

Relies on

as of 2026-11-13

Sources

  1. Does the Metal Matter? An overview of next-generation stent platforms. Cardiac Interventions Today, May/June 2010 (Cannon, Daniel, Hood and Yakubov), 2010, pp. 41 to 47; strut thickness text; Table 1, nominal elemental composition by weight
  2. State of the art: coronary artery stents – past, present and future EuroIntervention (Stefanini, Byrne, Windecker and Kastrati), 2017, EuroIntervention 2017;13:706-716
  3. Summary of Safety and Effectiveness Data, XIENCE V Everolimus Eluting Coronary Stent System (PMA P070015) US Food and Drug Administration (applicant Abbott Vascular, Cardiac Therapies), 2008, FDA notice of approval 2 July 2008; device description
  4. What Is Balloon Angioplasty and Stenting? Medtronic (UK patient site)
  5. Inception of the coronary stent: a story of successful collaboration between innovative scientists and the biotechnology industry REC: Interventional Cardiology 6(4):321-331 (Macaya-Ten, Gonzalo, Escaned, Macaya), 2024
  6. The Surprising History Behind an Extremely Common Cardiovascular Medical Device MD+DI, no byline or date shown on the page
  7. The new pricing policy for coronary stents in India: a boon or a bane? EuroIntervention (Upendra Kaul), 2017, EuroIntervention 2017;13(3), published 20 June 2017, DOI 10.4244/EIJV13I3A40
  8. Mission to provide Affordable, Quality Healthcare for All: Coronary Stent Prices capped, to bring down cost by about 380% Press Information Bureau, Government of India (Ministry of Chemicals and Fertilizers), 2017, posted 14 February 2017