Cardiac pacemaker: Interventions

Interventions

Changes that would cut emissions or material use, by lever and life stage, with effect per unit, readiness, adoption, evidence and who acts

LeverLife stageChangeEffect per unitReadinessAdoptionWho actsSource
CircularityEnd of lifeReuse explanted pacemakers that still have battery lifeOne reconditioned device replaces a new one; in a randomised trial of 306 patients, 12-month procedure-related infection was 1.6% for reconditioned and 3.1% for new devices [14]Piloted (randomised controlled trial in seven countries, ESC Congress 2025)Reuse is prohibited in the United States; re-sterilised devices may be exported for reuse [13]Regulators, hospitals, charities[14][13]
DesignComponents designLeadless design, with generator and electrode in one capsuleThe Micra weighs 1.75 g and fills 0.8 cc [2]; four BIOTRONIK Evia models weigh 24 to 26 g [1] (maker table values; different makers, no like-for-like comparison)Deployed (FDA approval April 2016 [6])Not given in any source openedDevice makers[2][1][6]
EfficiencyUseLonger battery life, so fewer replacement operationsReplacements were 264,824 of 1,002,664 implants in 2009 (26.4%, calculated) [4]; lithium iodine cells last about 10 years [3], the Micra AV2 is listed at 8 to 13 years [2]DeployedNot given in any source openedDevice makers[4][3][2]
Use lessUseRemote monitoring of pacemakers in place of clinic visitsA 2023 review says remote monitoring of pacemaker devices can reduce emissions; its Table 2 gives one study's 15 kg CO2e saved per remote monitoring cycle, a per-cycle figure from one Spanish programme, not a per-device figure [20]DeployedNot given in any source openedHospitals, device makers[20]

as of 2026-11-13

Sources

  1. BIOTRONIK EVIA User And Technical Manual BIOTRONIK (copy hosted on manualmachine.com), 2009, Doc. Id. 365353-A, cover dated 23.04.2009; mechanical characteristics, battery data and service time tables; explantation and disposal section
  2. Micra AV and Micra VR leadless pacemakers for bradyarrhythmia management Medtronic (India site), 2024, specification table; page footer says last updated March 2024
  3. A Look Back: 85 Years of Cardiac Pacing EP Lab Digest (HMP Global Learning Network), B. J. Wilhelm and B. Osswald, 2014, September 2014
  4. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009 - a World Society of Arrhythmia's project Pacing and Clinical Electrophysiology (Mond and Proclemer), on PubMed, 2011, abstract
  5. Premarket Approval (PMA): Medtronic Micra Transcatheter Pacemaker System, P150033 US Food and Drug Administration, Decision Date 04/06/2016; applicant Medtronic, Inc.
  6. Reconditioned pacemakers worked as well as new ones American Heart Association newsroom, 2024, dated 17 November 2024; the study abstract was not peer reviewed, as the page says
  7. Reconditioned pacemakers provide new hope for patients in low- and middle-income countries European Society of Cardiology (press release), 2025, dated 1 September 2025, ESC Congress
  8. Environmental impact of cardiovascular healthcare Open Heart (BMJ), Barratt and others, Open Heart 2023;10:e002279, 2023, abstract; Table 2 (Study findings), Perez Diaz et al. row