Material Insights · Understanding the Material World

Cardiac pacemaker: Working

What it is

The product and the job it does, in a short paragraph

A cardiac pacemaker is an implanted electronic device whose electrodes sense the heart's own beat and deliver electrical impulses when the heartbeat is slower than normal [11]. The functional unit is one pacemaker: a pulse generator with its leads, or a leadless capsule placed inside the heart. This node covers HS 9021.50. Defibrillators, leads sold alone, programmers and external temporary pacemakers are left out.

as of 2026-11-13

How it works

How it does its job, step by step in plain words

The pulse generator is a titanium case that holds a sealed battery, with an epoxy resin header where leads plug in [1]. Leads are wires that run through veins into the heart, and electrodes at their tips sense the natural heartbeat [11]. When the heartbeat is slower than normal, the generator sends an electrical impulse down the lead [11]. In a traditional device the generator sits in the chest or abdomen [11]. A leadless pacemaker puts generator and electrode in one capsule, about the size of a large pill, inside a heart chamber [11]. When the battery reaches its replacement time, the maker's manual says the pacemaker must be replaced [1]; about a quarter of the implants counted in 2009 were replacements (calculated) [4]. The working principle is on the twin technology page.

as of 2026-11-13

Key specifications

The numbers that describe it, each with unit, variant, year and source

SpecValueUnitVariantYearSource
Mass of the lightest of four BIOTRONIK Evia models (SR-T, single chamber), manual table value24gTransvenous2009[1]
Mass of the heaviest of four BIOTRONIK Evia models (DR, dual chamber), manual table value26gTransvenous2009[1]
Volume of the Evia DR (dual chamber), manual table value11cm3Transvenous2009[1]
Mass of a modern generator, about this much and the size of a matchbook (EP Lab Digest review)30gTransvenous2014[3]
Lead body diameter of one lead model (Boston Scientific INGEVITY+), spec sheet table value1.9mmTransvenous2026[10]
Typical life of a lithium iodine battery, about10yearsTransvenous2014[3]
Mass of the Micra AV2 and Micra VR3 leadless pacemakers, maker's specification table value1.75gLeadless2024[2]
Volume of the Micra AV2 and Micra VR3, maker's specification table value0.8cm3 (the page says cc)Leadless2024[2]
Battery longevity of the Micra AV2, low end (table value; 100% VDD pacing at 60 bpm)8yearsLeadless2024[2]
Battery longevity of the Micra AV2, high end (table value; 15% VDD pacing at 70 bpm)13yearsLeadless2024[2]

as of 2026-11-13

Variants

The variants whose make-up or footprint differs a lot, what defines each and its share of units

VariantDefining specYearSource
TransvenousGenerator in the chest or abdomen with one or more leads through veins into the heart; the four BIOTRONIK Evia models weigh 24 to 26 g and fill 10 to 12 cm3 (manual table values) [1]2009[1][11]
LeadlessGenerator and electrode in one capsule inside the heart; the Medtronic Micra weighs 1.75 g and fills 0.8 cc (maker's table values) [2]2024[2][11]

as of 2026-11-13

Limits

The physical limits set by science and the engineering limits of today's design, with today's value

LimitKindTodayUnitSource
Battery life of the lithium iodine cell before the generator is replaced, about (2014 review)Engineering10years[3]
Mass of a modern transvenous generator, about (2014 review)Engineering30g[3]
Mass of a leadless pacemaker (Micra, maker's table value)Engineering1.75g[2]

as of 2026-11-13

Science it relies on

The science principles it works by or is limited by

Relies on
  • Bioelectricity The electrodes deliver electrical impulses to the heart when the heartbeat is slower than normal [11].
  • The electrochemical cell The generator runs on a sealed lithium battery cell; the Evia manual lists lithium iodine among its battery systems [1][3].

as of 2026-11-13

Sources

  1. BIOTRONIK EVIA User And Technical Manual BIOTRONIK (copy hosted on manualmachine.com), 2009, Doc. Id. 365353-A, cover dated 23.04.2009; mechanical characteristics, battery data and service time tables; explantation and disposal section
  2. Micra AV and Micra VR leadless pacemakers for bradyarrhythmia management Medtronic (India site), 2024, specification table; page footer says last updated March 2024
  3. A Look Back: 85 Years of Cardiac Pacing EP Lab Digest (HMP Global Learning Network), B. J. Wilhelm and B. Osswald, 2014, September 2014
  4. The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009 - a World Society of Arrhythmia's project Pacing and Clinical Electrophysiology (Mond and Proclemer), on PubMed, 2011, abstract
  5. INGEVITY™+ Pacing Lead Boston Scientific, 2026, spec sheet CRM-2625410-AA, materials and dimensions table
  6. Pacemakers - How They Work National Heart, Lung, and Blood Institute (NIH), last updated 24 March 2022