CT scanner: Interventions
Interventions
Changes that would cut emissions or material use, by lever and life stage, with effect per unit, readiness, adoption, evidence and who acts
| Lever | Life stage | Change | Effect per unit | Readiness | Adoption | Who acts | Source |
|---|---|---|---|---|---|---|---|
| Efficiency | Use | Put the scanner in Off or low-power mode overnight instead of idle | COCIR's exemplary values are 72 kWh per day idle and 37 kWh per day Off; the 35 kWh difference (about 49%) is calculated [11] | Deployed (scenarios defined in the COCIR scheme) | Not stated | Hospitals and imaging centres | [11] |
| Efficiency | Use | Use an energy-savings mode | GE HealthCare says Energy Savings Mode cuts energy use by up to 15% on the Revolution Maxima [13] | Deployed | Not stated | Makers and hospitals | [13] |
| Design | Components design | Drop lead counterweights and cut lead in shielding | Siemens removed lead counterweights and says it cut lead shielding while protection stays equal [3]; GE HealthCare uses steel as counterweight [13]; no mass figure found | Deployed | Not stated | Makers | [3][13] |
| Circularity | End of life | Refurbish X-ray tube assemblies and reuse their parts | Up to 40% of a Siemens tube assembly may consist of reused parts; GE HealthCare reuses 60.1% of the mass of its tubes [3][13] | Deployed | Siemens takes tube assemblies back at end of life; no volume found | Makers | [3][13] |
| Circularity | Making | Recycle gadolinium oxide in detector ceramic production | About 25% of the gadolinium oxide processed each year comes from the closed loop, and the oxide supplied per area fell 82% against 2000/2001 [3] | Deployed | About 25% of annually processed gadolinium oxide [3] | Makers | [3] |
| Business model | End of life | Refurbish and resell whole scanners | Siemens refurbishes systems and reuses components and replacement parts where possible [3]; a sector report puts the refurbishment potential of CT at 85% (Table 6 value, not defined) [10]; no energy or emissions saving per unit was confirmed | Deployed (Siemens Lifecycle Product business [3]) | Not stated | Makers, service firms, hospitals | [3][10] |
as of 2026-11-14
Sources
- SOMATOM Pro.Pulse · Environmental Product Declaration (17893 0426) Siemens Healthineers AG, 2026, 'Climate change: total' chart; Key environmental performance indicators table
- Medical Imaging Equipment: State of Sustainability Research (final, 16 June 2022) Global Electronics Council for the Medical Equipment Proactive Alliance (MEPA), 2022, Tables 2 and 6
- COCIR Self-Regulatory Initiative for the Ecodesign of Medical Imaging Equipment: Status Report 2018 COCIR, 2018, Section 4.3.1
- Revolution Maxima: Focusing on sustainability in computed tomography (JB19858XX) GE HealthCare, 2024