CT scanner: Working

What it is

The product and the job it does, in a short paragraph

A CT scanner is a ring-shaped machine that turns an X-ray tube and a ring of detectors around a patient on a moving table and has a computer build slice images of the inside of the body. The functional unit is one complete medical CT scanner: gantry with tube and detectors, patient table, generator, cooling and operator console. Industrial and security CT, tubes and detectors sold alone, and plain X-ray machines (their own node) are left out.

as of 2026-11-14

How it works

How it does its job, step by step in plain words

The tube and the detectors sit on a rotating ring inside the gantry. The newest scanners spin it in as little as 0.33 s, and the table slides the patient through the ring [5]. Early scanners used a thin pencil beam, widened into a small fan in the second generation; in the third generation the tube and a detector array turn together around the patient, and that geometry is used in all commercial scanners today [1]. Detectors of the common kind turn X-rays into light in a ceramic and then into a signal. Photon-counting detectors convert each X-ray photon directly into an electrical signal [7]. Less than 1% of the tube's electron energy leaves as X-rays and more than 99% is heat, so tubes need strong cooling [30]. The computer turns thousands of readings into slices. The twin technology page explains the reconstruction.

as of 2026-11-14

Key specifications

The numbers that describe it, each with unit, variant, year and source

SpecValueUnitVariantYearSource
Weight of a SOMATOM Pro.Pulse CT scanner (Siemens declaration)2700kg (at most)Conventional2026[3]
Weight of a NAEOTOM Alpha.Pro CT scanner (Siemens declaration)3300kg (approx.)Photon-counting2025[4]
Fastest gantry rotation time (SOMATOM Pro.Pulse; also 0.5 s and 1.0 s)0.33sConventional2026[5]
Temporal resolution (SOMATOM Pro.Pulse)86msConventional2026[5]
Temporal resolution, best case (NAEOTOM Alpha.Pro)66ms (down to)Photon-counting2026[6]
Generator power (SOMATOM Pro.Pulse, 2 x 75 kW)150kWConventional2026[5]
Power of each of the two tubes (NAEOTOM Alpha.Pro)120kW per tubePhoton-counting2026[6]
Maximum power load (SOMATOM Pro.Pulse)200kVA (at most)Conventional2026[3]
Maximum power load (NAEOTOM Alpha.Pro)350kVA (at most)Photon-counting2025[4]
In-plane spatial resolution in Quantum HD mode (NAEOTOM Alpha.Pro)0.11mmPhoton-counting2026[6]

as of 2026-11-14

Variants

The variants whose make-up or footprint differs a lot, what defines each and its share of units

VariantDefining specYearSource
ConventionalScintillator detector: a gadolinium oxysulfide ceramic (UFC) turns X-rays into light first [9]; reference SOMATOM Pro.Pulse with two tubes and two detectors [5]2026[3][5][9]
Photon-countingPhoton-counting detector that converts each X-ray photon directly into an electrical signal; the first such scanner, the NAEOTOM Alpha, has a cadmium telluride layer [7]. Reference NAEOTOM Alpha.Pro with two tubes and two QuantaMax detectors [6], which Siemens describes as crystal semiconductors [32]; its declaration names no detector material [4]2025[4][6][7][32]

Indicative · as of 2026-11-14. Each variant is one Siemens model, the only ones with a published life-cycle declaration. Both are dual-source scanners, so weight and footprint also reflect that design. No source gives shares of units sold, and none opened names the detector material of the NAEOTOM Alpha.Pro itself.

Limits

The physical limits set by science and the engineering limits of today's design, with today's value

LimitKindTodayUnitSource
Share of the tube's electron energy that leaves as X-rays; more than 99% is heatPhysical1% (less than)[30]
Shortest time needed for one image, which sets how well a beating heart is frozen (best case today)Engineering66ms[6]
Finest detail resolved in the image plane in the best mode of today's photon-counting scannerEngineering0.11mm[6]

as of 2026-11-14

Science it relies on

The science principles it works by or is limited by

Relies on
  • X-rays Dense tissue absorbs more X-rays, so the beam that gets through carries the picture.
  • Fourier analysis Rebuilding a slice from many line measurements is a mathematical problem; the twin technology page explains it.
Limited by
  • X-rays X-rays ionise tissue, so EU rules require each dose to fit the medical purpose of the exposure [26]; regulators also license the equipment [17].

as of 2026-11-14

Sources

  1. CT Archives - AAPM Virtual Museum AAPM Virtual Museum (American Association of Physicists in Medicine)
  2. SOMATOM Pro.Pulse · Environmental Product Declaration (17893 0426) Siemens Healthineers AG, 2026, 'Climate change: total' chart; Key environmental performance indicators table
  3. NAEOTOM Alpha.Pro · Environmental Product Declaration (17247 1125) Siemens Healthineers AG, 2025, 'Climate change: total' chart; Key environmental performance indicators table
  4. SOMATOM Pro.Pulse Dual Source CT scanner Siemens Healthineers, 2026
  5. NAEOTOM Alpha.Pro Siemens Medical Solutions USA, 2026
  6. Siemens Healthineers Announces FDA Clearance of NAEOTOM Alpha, World's First Photon-Counting CT Siemens Healthineers, 2021
  7. UFC - Ultra Fast Ceramic Siemens Healthineers, 2026
  8. AERB Annual Report 2023, Chapter 2: Safety Surveillance of Radiation Facilities Atomic Energy Regulatory Board, India, 2023, Tables 2.1 and 2.2
  9. Council Directive 2013/59/Euratom of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation legislation.gov.uk (EU text as it stood on 31 December 2020), 2013, Article 5
  10. X-ray production Radiopaedia.org
  11. Photon-counting CT by Siemens Healthineers: Quantum Technology Siemens Healthineers, 2026